Calibration Laboratory
Kinetics Corporation Ltd. has established the calibration lab in Thailand since 1994. The scope of calibration lab covers the area of all pressure device including dead weight testers and pressure gauges in hydraulic and pneumatic. Our lab calibration expertise passed through all the in house and outsource training. We have all the reliable equipments from the manufacturer DH-Budenberg, UK, certified to the international standard of UKAS. Our laboratory has passed the crucial inspection of ISO/IEC 17025 which is the industrial standard of calibration lab. In the near future, we are going to expand the lab network to Temperature, Electrical and Velocity for one stop visit and provide the best service support to our valuable customer.

Capability of Calibration Laboratory
| Field of Measurement | Parameter/Range/Item |
Calibration and Measurement Capability* |
Standard/Technique/Method/Equipment/Remark |
|---|---|---|---|
|
Pressure |
Positive pressure , hydraulic Dead weight tester 100 kPa to 6000 kPa (10 psi to 800 psi) |
10 x 10-5 x Pe but not less than 0.06 kPa |
In-house method: WI01-PL based on DKD R6-1 Fluid Medium : Nitrogen Gas (N2) |
|
Pressure
|
20 bar to 1200 bar (200 psi to 16000 psi) |
10 x 10-5 x Pe but not less than 1.20 kPa |
|
| Pressure |
2 bar to 300 bar (30 psi to 4000 psi) |
10 x 10-5 x Pe but not less than 0.30 kPa |
|
|
Pressure
|
Pressure measuring instrument 0 bar to 60 bar (0 psi to 800 psi) |
25 x 10-5 x Pe but not less than 0.15 kPa |
In-house method: WI01-PL based on DKD R6-1 Fluid Medium : Oil (ISO VG22) |
|
Pressure
|
0 bar to 1200 bar (0 psi to 16000 psi) |
25 x 10-5 x Pe but not less than 3.00 kPa |
|
|
Pressure
|
0 bar to 300 bar (0 psi to 4000 psi) |
25 x 10-5 x Pe but not less than 0.75 kPa |
|
|
Pressure
|
Positive pressure, pneumatic Dead weight tester 15 mbar to 1000 mbar (5 in.water to 400 in.water) |
10 x 10-5 x Pe but not less than 1.00 Pa |
In-house method: WI03-PL based on EURAMET cg-3 (Method A) Fluid Medium : Nitrogen Gas (N2) |
|
Pressure
|
0.2 bar to 25 bar (3 psi to 400 psi) |
10 x 10-5 x Pe but not less than 0.03 kPa |
|
|
Pressure
|
1 bar to 120 bar (10 psi to 1600 psi) |
10 x 10-5 x Pe but not less than 0.12 kPa |
***expressed as an uncertainty (±) which for k = 2, providing a level of approximately 95%
Capability of Calibration Laboratory
| Field of Measurement | Parameter/Range/Item |
Calibration and |
Standard/Technique |
|---|---|---|---|
|
Pressure
|
Pressure Measuring Instrument 0 mbar to 1000 mbar (5 in.water to 400 in.water) |
25 x 10-5 x Pe but not less than 2.50 Pa |
In-house method: WI01-PL based on DKD R6-1 Fluid Medium : Nitrogen Gas (N2) |
|
Pressure
|
0 bar to 25 bar (0 psi to 400 psi) |
25 x 10-5 x Pe but not less than 0.06 kPa |
|
|
Pressure
|
0 bar to 120 bar (0 psi to 1600 psi) |
25 x 10-5 x Pe but not less than 0.30 kPa |
|
|
Pressure
|
Negative pressure, pneumatic Pressure Measuring Instrument 0 mbar to 1000 mbar (5 in.water to 400 in.water) |
25 x 10-5 x Pe but not less than 2.50 Pa |
In-house method: WI01-PL based on DKD R6-1 Fluid Medium : Nitrogen Gas (N2) |
***expressed as an uncertainty (±) which for k = 2, providing a level of approximately 95%
Kinetics Corporation Ltd. (Calibration Laboratory)
388 Ratchadapisek Road, Chandrakasem,
Chatuchak, Bangkok 10900 Thailand.
Tel : 662-515-8951
Fax : 662-515-8988
E-mail : calibration_tib@kinetics.co.th